EPA Method 537.1 Instrumentation Guide

EPA Method 537.1 determines 18 selected per- and polyfluoroalkyl substances (PFAS) in drinking water. A 250 mL sample is extracted on a polystyrenedivinylbenzene (SDVB) solid phase extraction (SPE) cartridge, concentrated to 1 mL, and analyzed by LC-MS/MS.

A photo of a refurbished AGILENT LC-MS system for EPA METHOD 537.1 that links to a catalog of Liquid Chromatography-Mass Spectrometry equipment.

Single-laboratory lowest concentration minimum reporting levels (LCMRLs) in the method range from 0.53 to 6.3 ng/L. A lab’s actual reporting levels depend on its instrument, background contamination, and calibration.

Read EPA Method 537.1, Version 2.0.

Looking for an LC-MS/MS system for EPA Method 537.1? Browse refurbished LC/MS systems or request a configuration quote.

EPA Method 537.1 Summary

Item

Detail

Method

EPA Method 537.1, Version 2.0, March 2020 (EPA/600/R-20/006)

Analytes

18 PFAS: HFPO-DA, NEtFOSAA, NMeFOSAA, PFBS, PFDA, PFDoA, PFHpA, PFHxS, PFHxA, PFNA, PFOS, PFOA, PFTA, PFTrDA, PFUnA, 11Cl-PF3OUdS, 9Cl-PF3ONS, ADONA

Matrix

Drinking water

Technique

SPE with LC-MS/MS

Sample

250 mL

SPE cartridge

0.5 g, 6 mL SDVB

Final extract

1 mL in 96:4 methanol:water

Injection

10 µL used in method development

Ionization

Negative ion electrospray (ESI)

LCMRLs

0.53 to 6.3 ng/L, depending on analyte

Holding times

Extract samples within 14 days; analyze extracts within 28 days

Method 537.1 applies to drinking water. Other sample types need a different method. EPA Method 1633 covers matrices such as wastewater, soil, and tissue.

Instrument requirements

The LC must inject up to 10 µL reproducibly and run binary linear gradients at a constant flow rate near 0.3 mL/min. The LC-MS/MS must perform negative ion electrospray at that flow rate and produce unique product ions within set retention time segments. The method requires at least 10 scans across each chromatographic peak.

The data system must integrate ion abundance, calculate response factors, and build linear or quadratic calibration curves. The method names no required LC-MS/MS brand. Its validation data were generated on a Waters XEVO TQMS triple quadrupole.

Setting

Method development condition

Column

C18, 2.1 × 150 mm, 5 µm (Waters Atlantis dC18 or equivalent)

Mobile phase A

20 mM ammonium acetate

Mobile phase B

Methanol

Gradient

40% to 90% methanol over 25 min, with a total program of 37 min

Flow rate

0.3 mL/min

Injection

10 µL

Polarity

Negative ion

Capillary voltage

-3 kV

Desolvation temperature

350 °C

Collision gas

Argon; nitrogen allowed with sufficient sensitivity

Labs may change the LC column, mobile phase composition, LC conditions, and MS and MS/MS conditions. Each change requires repeating the initial demonstration of capability. Sample collection, preservation, extraction steps, and QC requirements cannot be changed.

PFOS, PFBS, and PFHxS must be quantified with the precursor to m/z 80 transition. A system that cannot measure this transition cannot be used for those three analytes.

Know the configuration you need? Request an EPA 537.1 LC-MS/MS quote.

What to look for in a refurbished LC-MS/MS system

The right refurbished system combines the required capabilities with evidence that it performs reliably. Use these checks when comparing systems.

System area

Method 537.1 need

What to check

Mass spectrometer

MS/MS with negative ion ESI

Negative ion performance and source condition

Required transitions

Precursor to m/z 80 for PFOS, PFBS, and PFHxS

Demonstrated response for these transitions

Scan speed

At least 10 scans across each peak

Peak sampling at your planned run time

LC pump

Binary linear gradients near 0.3 mL/min

Gradient accuracy and flow stability

Autosampler

Reproducible injections up to 10 µL

Injection repeatability and polypropylene vial compatibility

PFAS background

Reagent blanks below 1/3 of the MRL

PTFE in the flow path, PEEK replacements, and any delay column

Software

Peak integration and calibration

Included workstation and data system

Documentation

The lab must prove system performance

Test reports, calibration data, and configuration records

EPA found that PFAS built up in PTFE solvent lines during idle periods. The method’s developers replaced those lines with PEEK tubing and replaced PTFE solvent frits with stainless steel frits.

SPE equipment is also part of the workflow. Plan for a vacuum manifold or an automated SPE system, along with a nitrogen evaporation system.

Need help comparing systems? Tell our team about your sample volume, reporting levels, and preferred platform. Talk to an instrument specialist or call (844) AMP-TECH.

Example LC-MS/MS configurations

The platforms below are examples that may be configured for EPA Method 537.1. The method names no required brand or model, so your choice can reflect budget, platform preference, and service needs.

Platform

Configuration for Method 537.1

Key Benefits for EPA 537.1

Agilent Ultivo

1260 Infinity II binary pump, Vialsampler, multicolumn thermostat, PFAS background controls

Agilent has published second-lab data for EPA 537.1 on the Ultivo using a 1290 Infinity II LC.

Waters Xevo TQ-S micro

ACQUITY I-Class binary solvent manager, FTN sample manager, column manager, PFAS background controls

Waters has published EPA 537.1 work on this mass spectrometer with an I-Class PLUS and PFAS kit.

Agilent 6460

1260 Infinity binary pump, ALS autosampler, column compartment, PFAS background controls

A complete triple quadrupole system with Agilent Jet Stream ESI and a mass range of m/z 5 to 3,000.

Agilent 6490 with 1260 Infinity II

1260 Infinity II binary pump, Vialsampler, PFAS background controls

A higher-sensitivity iFunnel platform for labs targeting low reporting levels. The DAD is not required by the method.

Agilent 6490 with 1200/1260

1260 Infinity binary pump, 1200 autosampler, column compartment, PFAS background controls

An iFunnel triple quadrupole with a complete binary LC stack. The VWD is not required by the method.

Thermo Scientific TSQ Quantis

Vanquish binary pump, split sampler, column compartment, HESI source, PFAS background controls

Includes a binary UHPLC stack and a mass range of m/z 5 to 3,000. The VWD is not required by the method.

Agilent 6490 mass spectrometer

Pair with a compatible binary LC, PFAS background controls

A mass spectrometer-only option for labs that already have a compatible Agilent LC.

Confirm each system’s negative ion performance, m/z 80 response, and PFAS background during configuration. Our team can review LC hardware, delay column options, and your target reporting levels before you buy.

Don’t see the right configuration in the store? Tell us what you need, and we’ll check the available options.

Choosing a configuration

Start with your target reporting levels, sample load, and existing LC.

  • Published method data: compare the Agilent Ultivo and Waters Xevo TQ-S micro if vendor-published EPA 537.1 work matters to your team.
  • Lower capital cost: compare complete systems such as the Agilent 6460 with 1260 HPLC.
  • Lower reporting targets: consider a higher-sensitivity platform, then verify your MRLs and system background.
  • Existing LC: consider a mass spectrometer-only option after confirming LC compatibility and PFAS background controls.
  • Compliance monitoring: confirm laboratory certification and reporting requirements with your EPA or state program.
  • Other drinking water PFAS: review EPA Method 533 if you need additional shorter-chain compounds.

Amptius Certified Refurbished Instruments

AMP Tech offers refurbished analytical instruments and can help you configure a refurbished LC-MS/MS system for EPA Method 537.1. Amptius Certified refurbished instruments go through a six-stage certification process, including repair, analytical testing, and documentation before shipment, giving your lab a lower-cost alternative to buying new.

Certified instruments include a 6-month limited warranty covering parts and labor for covered defects in materials or workmanship.

AMP Guard Service Plans

For support beyond the warranty, AMP Guard offers optional service plans with remote support, service facility repair, and on-site options. Coverage depends on the plan and instrument.

Request a quote with your sample volume, reporting levels, and preferred platform. Our documentation and technical support can help your team prepare for method validation and ongoing QC.

Sell or Trade in your lab equipment

Upgrading your lab or clearing surplus equipment? AMP Tech buys GC/MS, LC/MS and triple quad, HPLC, ICP-OES and ICP-MS, and FTIR and spectroscopy systems, along with supporting lab equipment. We buy everything from a single instrument to an entire lab, anywhere in the country.

You get a firm, no-obligation offer within 48 hours with no fees or commission. Our team handles decommissioning, packing, and removal, and you can apply your equipment’s value toward an Amptius Certified instrument.

Sell your equipment or check your trade-in value.

Sources

Ready to Configure Your EPA 537.1 LC-MS/MS System?