EPA Method 8011 Instrumentation Guide

EPA Method 8011 determines 1,2-dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP) in drinking water and groundwater. A 35 mL sample is extracted with 2 mL of hexane, and 2 µL of the extract is injected into a gas chromatograph with a linearized electron capture detector (GC-ECD).

A photo of AGILENT GC-ECD System for EPA Method 8011 that links to a chatalog of Gas Chromatography equipment.

The method has a working range of about 0.03 to 200 µg/L and a reported method detection limit (MDL) of 0.01 µg/L for each analyte. Actual detection limits depend on the GC system, sample matrix, and calibration.

Read EPA SW-846 Method 8011.

Looking for a GC-ECD system for EPA Method 8011? Browse refurbished GC systems or request a configuration quote.

EPA Method 8011 Summary

Item Detail

Method

SW-846 Method 8011, Revision 0, July 1992

Analytes

EDB (CAS 106-93-4); DBCP (CAS 96-12-8)

Matrices

Drinking water and groundwater
Technique

Microextraction with GC-ECD

Sample and solvent

35 mL sample; 2 mL hexane

Injection

2.0 µL

Working range

About 0.03 to 200 µg/L

MDL

0.01 µg/L for each analyte

Extraction and analysis time

30 to 50 minutes per sample

Column 1 Value 9 Column 2 Value 9

Other compounds or matrices require precision and accuracy data on actual samples, plus qualitative confirmation by GC/MS.

Instrument requirements

The GC must be capable of temperature programming. It should have a linearized ECD and a capillary column splitless injector. An autosampler is strongly recommended, although manual injection with the solvent flush technique is allowed.

Setting

Column A (primary)

Column B (confirmation)

Column

30 m × 0.32 mm ID fused silica, 0.25 µm film

30 m × 0.32 mm ID fused silica, 0.25 µm film

Phase

Durawax-DX 3 or equivalent

DB-1 or equivalent

Injector

200 °C

200 °C

Detector

290 °C

290 °C

Carrier gas

Helium, 25 cm/sec

Helium, 25 cm/sec

Initial oven

40 °C, hold 4 min

40 °C, hold 4 min

Ramp

To 190 °C at 8 °C/min

To 270 °C at 10 °C/min

Final hold

25 min, or until analytes elute

10 min, or until analytes elute

EDB retention time

9.5 min

8.9 min

DBCP retention time

17.3 min

15.0 min

Column A is the primary analytical column. Column B provides confirmation when GC/MS is not available.

Know the configuration you need? Request an EPA 8011 GC-ECD quote.

What to look for in a refurbished GC-ECD system

The right refurbished system combines the required hardware with evidence that it performs reliably. Use these checks when comparing systems.

System area

Method 8011 need

What to check

GC and oven

Temperature programming from 40 °C up to 270 °C for Column B

Oven control, temperature ramps, and retention time stability

Inlet

Capillary column splitless injection at 200 °C

Splitless capability and leak-tight pneumatics

Detector

Linearized ECD at 290 °C

ECD condition, licensing, and transfer requirements

Autosampler

Automated 2.0 µL injections strongly recommended

Injection repeatability and compatibility with the GC

Columns

Primary and confirmation columns

Whether Column A, Column B, or a confirmed equivalent is included

Carrier gas

Helium at 25 cm/sec

Gas connections and flow control

Software

Calibration, peak response, and reporting

Included workstation and data system

Documentation

The lab must prove system performance

Test reports, calibration data, and configuration records

Need help comparing systems? Tell our team about your sample volume, confirmation approach, and preferred platform. Talk to an instrument specialist or call (844) AMP-TECH.

Example GC-ECD configurations

The platforms below are examples that may be configured for EPA Method 8011. The method names no required GC brand or model, so your choice can reflect budget, platform preference, and service needs.

Platform

Configuration for Method 8011

Key Benefits for EPA 8011

Agilent 8890

Splitless-capable inlet, linearized ECD, 7650A autosampler

A modern platform with electronic pneumatic control and automated injection.

Agilent 8860

Split/splitless inlet, linearized ECD, 7693 autosampler

Dual split/splitless inlets can support separate primary and confirmation channels when configured with two ECDs.

Agilent 7890B

Splitless-capable inlet, linearized ECD, 7693 autosampler

Supports multiple oven ramps, well beyond the single ramp used in Method 8011, and the 7693 autosampler supports automated injections.

Agilent 7820A

Splitless-capable inlet, linearized ECD, 7693 autosampler

Agilent lists a micro-ECD option, and its oven ramp capability exceeds the method’s 8–10 °C/min programs.

Agilent 7890A

Splitless-capable inlet, linearized ECD, 7693 autosampler

Electronic pneumatic control and precise oven ramping support repeatable runs. The listed base system can be configured with a detector.

Agilent 6890N

Split/splitless inlet, linearized ECD, 7683B autosampler

A lower-cost starting point with an autosampler and split/splitless inlet.

Some listed systems use other detectors, but our team can help you plan an EPA Method 8011 configuration. We’ll review ECD options, columns, helium requirements, and your confirmation approach before you buy.

Don’t see the right configuration in the store? Tell us what you need, and we’ll check the available options.

Choosing a configuration

Start with how your lab will inject samples and confirm results.

  • Routine batches: choose an autosampler for 2.0 µL injections, as the method strongly recommends.
  • Column confirmation: plan for both the primary and confirmation columns.
  • GC/MS confirmation: a GC/MS can replace Column B confirmation, but the primary analysis remains GC-ECD. Browse refurbished GC/MS systems.
  • Other compounds or matrices: plan for additional precision, accuracy, and GC/MS confirmation data.

Amptius Certified Refurbished Instruments

AMP Tech offers refurbished analytical instruments and can help you configure a refurbished GC-ECD system for EPA Method 8011. Amptius Certified refurbished instruments go through a six-stage certification process, including repair, analytical testing, and documentation before shipment, giving your lab a lower-cost alternative to buying new.

Certified instruments include a 6-month limited warranty covering parts and labor for covered defects in materials or workmanship.

AMP Guard Service Plans

For support beyond the warranty, AMP Guard offers optional service plans with remote support, service facility repair, and on-site options. Coverage depends on the plan and instrument.

Request a quote with your sample volume, confirmation approach, and preferred platform. Our documentation and technical support can help your team prepare for method validation and ongoing QC.